Supplied with Melting Capillary one end sealed (Pack of 100 Nos) , Power Cord and Operational & Instruction Manual and Warranty Card.
Technical Specification :
- Detection Principal : Automatic by photocell
- Temperature Range : 5ºC Above Ambient to 400ºC
- Heating before set point : Automatic Temperature Heating depending upon temp. diff.
- Heating after Set point : Variable 0.5, 1.0 & 2.0 ºC – Selectable
- Detection of Melting : a) Melting Point b) Melting Range
- Oven Controller : Microcontroller Based
- Temperature Resolution : 0.1ºC
- Calibration : Multi Point Calibration Standard – as per Pharmacopia or Any users known standard.
- Max. Heating Time : Less then 10 Minutes from 50ºC to 350ºC
- Max. Cooling Time : Less then 10 Minutes from 350ºC to 50ºC
- Melting Capillary : One Capillary (Single Channel)
- Accuracy of Temp.:
- Ambient +10ºC to 100ºC………… ± 0.2ºC.
- 101ºC to 200ºC …………. ± 0.4ºC.
- 200ºto 350ºC …………….. ± 0.5ºC.
- Sample : Borosilicate Capillary Tube One End Sealed
- OD : 1.4 to 1.6 mm
- ID : 1.0mm
- Length : 100mm
- Filling Height : 3 mm
- Temperature Sensor : PT-100
- Data Storage : Non Volatile Memory with More than 50 analysis, data storage with corresponding calibration data.
- Method Storage : Ten User method parameter with Add/Edit Facility.
- Printer Output : Parallel Port (Optional)
- Keyboard : Alphanumeric Polyester Softkeys
- Backlit LCD Display : 20 x 4 Line Alphanumeric
- Power : 230 VAC ± 10%, 50Hz